Staphylococcus epidermidis Bacteria Reduction

Authors
R. Wormald
Facility
University of Huddersfield; England, UK
Download
Full Report

Objective

To evaluate the ability of the Protect 800/900 to remove Staphylococcus epidermidis aerosols.

Methodology

A 1.0 m3 air tight perspex chamber was used to evaluate the ability of the Protect 800/900 to remove Staphylococcus epidermidis aerosols. The chamber was fitted with an internal fan to maintain mixing, sampling and injection ports, and the sanitation unit. The fan and the sanitation unit were activated from outside of the chamber as and when required.

Summary of Results

The presence of the Protect 800/900 appeared to have significant impact on the levels of bacterial aerosols present in the sampling chamber over and above that seen in the control run. Both the rate of removal and the final log reduction are greater in the presence of the Protect 800/900. In over 30 minutes of sampling, the sanitation unit removed 95% of the contamination.

These results indicate that the device is able to make a significant impact on the levels of aerosol contamination under laboratory conditions. The levels of contamination used to challenge the system in this case are much higher than would be experienced in healthcare settings, suggesting that the Protect 800/900 is likely to be able to make a significant impact on airborne contamination.