Acinetobacter baumannii Bacteria Reduction
- Authors
- Alan Goldtein, Laboratory Manager
- Facility
- EMSL Analytic, Inc. - Morrisville, NC
- Download
- Full Report
Objective
To evaluate the efficacy and the dose response of NuvaWave Instant UV handheld device in reducing Acinetobacter baumannii bacteria on surfaces.
Methodology
Utilizing ASTM E3135 standard practice to evaluate the ultraviolet germicidal irradiation designed to kill/inactivate microorganisms deposited on inanimate carriers, such as steel or glass. The carriers were inserted into a computer controlled robotic NuvaWave test device (12 mW/cm2 ) irradiated the pathogen carriers at a separation distance of 2 inches for a defined period of time. The dose and time for each period was recorded. The irradiated pathogens were compared to a non-irradiated control group to determine the amount of inactivation.
Summary of Results
The NuvaWave handheld device reduced 99.93% of Acinetobacter baumannii after one second of exposure.